CFN: AP-XPS + IRRAS on one sample, UHV to ~20 mbar. PM-IRRAS on the Vertex 80v
Transmission
Beam through a pressed wafer, or powder held on a metal mesh.
+ Beer-Lambert holds: the one to use for counting sites.
− Opaque below about 1100 cm⁻¹ on oxides. Mass transfer.
CFN: Vertex 80v, powder on a tungsten mesh, UHV or gas, down to −120 °C
O-PTIR + Raman
optical photothermal IR
IR laser heats where the sample absorbs, a visible beam reads it out.
+ Sub-micron IR maps, Raman from the same spot.
− Laser range excludes the CO stretch. Not for monolayers.
CFN: mIRage, 758 to 1806 cm⁻¹, < 500 nm. In situ in development
DRIFTS
diffuse reflectance
Loose powder in a flow cell: the operando workhorse.
+ Real powders, no pressing, easy to couple with XAS or MS.
− Quantification is model-dependent. Bed gradients.
Not at the CFN. Elsewhere at BNL: Chemistry, some NSLS-II beamlines
O-PTIR: infrared chemistry at visible-light resolution
A conventional IR microscope cannot resolve anything smaller than several microns. O-PTIR reads the IR absorption with a visible beam, so resolution stops depending on the IR wavelength.
Which technique reaches which bonds?
Infrared on supported catalysts goes blind below about 1000 cm⁻¹, exactly where metal–carbon, metal–oxygen and metal–hydrogen bonds vibrate. Raman and HREELS reach them.
CO: one probe molecule, many answers
More back-donation from the metal into CO 2π* weakens the C–O bond: lower wavenumber. Position reports the site. Cations push it above the gas value.
Block 2
From raw data to numbers: the workflow
STEP 1RatioSample single beam over background. Choose the background on purpose.
STEP 2Remove the gasSubtract gas-phase and vapour references, scaled where no adsorbate absorbs.
STEP 3BaselineAnchor only where nothing absorbs. Same recipe for every spectrum.
STEP 4FitFewest components that leave a noise-like residual. Each one needs a name.
STEP 5Follow in timeBand areas against time, temperature or gas composition.
Every step is a model. Write down what you did, apply it identically to the whole series, and keep the raw single beams.
You just heard the same message for XPS. It is the same disease: a fit will always converge, and the software will never tell you it is meaningless.
Demonstration: an operando dataset, start to finish
Simulated: CO adsorption, then CO oxidation light-off, then purge, on a supported Pt catalyst. 30 spectra, one per minute.
Assignment: a band position is a hypothesis
A lookup table gives you candidates. These give you evidence: isotope labelling, reference compounds, coverage and temperature dependence, calculation.
Is band intensity proportional to coverage?
In DRIFTS, neither Kubelka-Munk nor log(1/R) is linear in general. And on metals, neighbouring CO molecules share intensity.
IRRAS on flat model catalysts: the surface selection rule
Only the dipole component perpendicular to the metal is seen, only with p-polarized light, best near grazing incidence. A missing band can mean a flat-lying molecule.
Spectator or intermediate?
A band that is present under reaction conditions is not therefore part of the mechanism. The largest bands usually belong to the most stable species, which are the least reactive ones.
Transient experiments: switch the gas, compare how fast the band decays with how fast the product forms.
Isotope switching (SSITKA): ¹²CO to ¹³CO at steady state. An intermediate exchanges at the rate of product labelling.
Modulation excitation: periodic perturbation plus phase-sensitive detection picks out species that respond, and rejects the static ones.
From the demo
Adsorbed CO: falls when CO₂ rises. Kinetically relevant.
Carbonate / formate: grows regardless. On the support, watching.
The second set gave the biggest bands in the fingerprint region.
A real example
Placeholder. Drop one of your own datasets here: raw single beams, the processed spectra, and the one number that came out of it. Put the image in slides/img/ and replace this block. Or delete the slide.
Pitfall checklist
Background taken at a different temperature than the sample spectrum
Gas reference recorded at a different resolution, pressure or temperature
Baseline anchors that move from spectrum to spectrum
More fit components than you can name
Reading a 10 cm⁻¹ shift as an oxidation-state change without checking coverage
Quoting ratios of band areas as ratios of sites
Trusting absorbance above about 1: too little light is left and the scale stops being linear
Assigning from position alone, with no isotope or reference check
Calling the biggest band the intermediate
Throwing away the single beams and keeping only processed spectra
Take home
Three things to remember
1. A spectrum is a ratio, and every processing step is a model. Record the recipe and apply it to the whole series.
2. Position suggests, isotopes prove. Coverage moves bands as much as chemistry does.
3. Present under reaction is not the same as involved in reaction. Test it with a transient.
All tools: catir.upton.club · open, free, no install.