Infrared Reflection–Absorption Spectroscopy
IRRAS
Also called RAIRS or IRAS
Adsorbates on flat, reflective surfaces — model catalysts and single crystals.
At BNL · Two in situ instruments at the CFN.
How it works
IR light reflects off a metal surface at grazing incidence (~80–85°). Only the p-polarized component builds a strong electric field normal to the surface, so only vibrations with a dipole component perpendicular to the surface are seen — the surface selection rule.
Geometry & samples
Grazing-incidence reflection from a flat, conductive substrate.
Polished metal single crystals, oxide thin films on metal, model catalysts.
Best for
- —Single crystals and thin films
- —Molecular orientation of adsorbates
- —UHV and near-ambient pressure studies (PM-IRRAS)
Strengths
- —Sub-monolayer sensitivity
- —Orientation information via selection rule
- —Clean, well-defined surfaces
Limitations
- —Requires flat, reflective substrate
- —Gas-phase interference at high pressure (use PM-IRRAS)
- —Weak signals on non-metals
Practical tips
- —Use p-polarized light and grazing angles
- —Polarization modulation subtracts gas-phase bands
- —CO is an excellent probe of site types (atop, bridge, hollow)
Instruments at the CFN
Ambient Pressure PES and IRRAS
AP-XPS and IRRAS chambers connected through vacuum, so the same sample is measured by both. IRRAS from UHV to about 20 mbar, sample from −180 °C to 500 °C.
CFN equipment page ↗Bruker Vertex 80v with PM-IRRAS chamber
Vacuum spectrometer feeding a polarization-modulated IRRAS chamber. Heating and cooling under UHV and gas.
CFN equipment page ↗