Optical Photothermal Infrared + Raman Microscopy
O-PTIR / Raman
Optical photothermal infrared spectroscopy
Infrared chemistry at sub-micron resolution, with Raman from the same spot.
At BNL · mIRage instrument at the CFN. In situ capabilities are being developed.
How it works
A pulsed, tunable IR laser is focused on the sample. Where the sample absorbs, it heats and expands slightly. A visible probe laser focused on the same spot detects that photothermal response. Sweeping the IR laser gives a spectrum that resembles a transmission IR spectrum, while the spatial resolution is set by the visible probe (about half a micron), not by the IR wavelength. The same probe beam also generates a Raman spectrum.
Geometry & samples
Co-focused IR pump and visible probe in a reflection microscope. Non-contact.
Powders dispersed on a substrate, films, cross-sections, pellets. Little or no preparation.
Best for
- —Mapping chemical heterogeneity between and within catalyst particles
- —Complementary IR and Raman from the same location
- —Samples too thick or too rough for transmission
Strengths
- —Sub-micron spatial resolution across the whole IR range
- —No contact with the sample, unlike AFM-IR
- —Spectra largely free of the scattering artefacts of IR microscopy
Limitations
- —The CFN instrument covers 758 to 1806 cm⁻¹: no C–H or O–H stretches, and no carbonyl CO region (1900 to 2300 cm⁻¹)
- —Signal depends on thermal properties, so intensities are not simply absorbance
- —Not a sub-monolayer adsorbate technique
Practical tips
- —Plan around the fingerprint region: carbonates, carboxylates, formates, C=O, framework and M=O bands are in range
- —Keep probe power low: dark catalysts burn easily
- —Use Raman for the oxide lattice and carbon, IR for the organics and adsorbates