Transmission Infrared Spectroscopy
Transmission IR
Quantitative absorbance through self-supported wafers.
At BNL · In situ at the CFN, on the Bruker Vertex 80v.
How it works
IR light passes straight through a thin, pressed catalyst wafer. Absorbance follows Beer–Lambert, so with known extinction coefficients you can count sites — e.g. Brønsted vs Lewis acid sites with pyridine.
Geometry & samples
Beam passes through a self-supporting pressed wafer in a vacuum/gas cell.
Self-supporting wafers, ~10–20 mg/cm², pressed from powder.
Best for
- —Quantitative site counting
- —Acidity with probe molecules (pyridine, NH₃, CO)
- —Zeolites and oxides
Strengths
- —Quantitative (Beer–Lambert)
- —Excellent signal-to-noise
- —Well-established extinction coefficients
Limitations
- —Sample must be IR-transparent enough
- —Pressing can alter the material
- —Opaque below ~1200 cm⁻¹ for many oxides
Practical tips
- —Normalize by wafer mass per area
- —Activate in vacuum before probe adsorption
- —Pyridine: ~1545 cm⁻¹ (Brønsted), ~1450 cm⁻¹ (Lewis)