Transmission Infrared Spectroscopy

Transmission IR

Quantitative absorbance through self-supported wafers.

At BNL · In situ at the CFN, on the Bruker Vertex 80v.

How it works

IR light passes straight through a thin, pressed catalyst wafer. Absorbance follows Beer–Lambert, so with known extinction coefficients you can count sites — e.g. Brønsted vs Lewis acid sites with pyridine.

Geometry & samples

Beam passes through a self-supporting pressed wafer in a vacuum/gas cell.

Self-supporting wafers, ~10–20 mg/cm², pressed from powder.

Best for

  • —Quantitative site counting
  • —Acidity with probe molecules (pyridine, NH₃, CO)
  • —Zeolites and oxides

Strengths

  • —Quantitative (Beer–Lambert)
  • —Excellent signal-to-noise
  • —Well-established extinction coefficients

Limitations

  • —Sample must be IR-transparent enough
  • —Pressing can alter the material
  • —Opaque below ~1200 cm⁻¹ for many oxides

Practical tips

  • —Normalize by wafer mass per area
  • —Activate in vacuum before probe adsorption
  • —Pyridine: ~1545 cm⁻¹ (Brønsted), ~1450 cm⁻¹ (Lewis)